MPS-65-15

Description
65W Single Output Medical Type
File Size338KB,2 Pages
ManufacturerTotal Power International
Websitehttp://www.total-power.com/
Download Datasheet Compare View All

MPS-65-15 Online Shopping

Suppliers Part Number Price MOQ In stock  
MPS-65-15 - - View Buy Now

MPS-65-15 Overview

65W Single Output Medical Type

65W Single Output Medical Type
Features :
MPS-65
Universal AC input / Full range
Low leakage current
0.3mA
Protections:Short circuit/Over load/Over voltage
Cooling by free air convection
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
series
SPECIFICATION
MODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER
OUTPUT
OUTPUT POWER (max.)
VOLTAGE ADJ. RANGE
MPS-65-3.3
3.3V
12A
0 ~ 15.2A
39.6W
MPS-65-5
5V
12A
0 ~ 13.8A
60W
100mVp-p
MPS-65-7.5
7.5V
8A
0 ~ 9.6A
60W
100mVp-p
MPS-65-12
12V
5.2A
0 ~ 6A
62.4W
100mVp-p
MPS-65-13.5 MPS-65-15
13.5V
4.7A
0 ~ 5.4A
63.5W
100mVp-p
15V
4.2A
0 ~ 4.8A
63W
100mVp-p
MPS-65-24
24V
2.7A
0 ~ 3A
64.8W
100mVp-p
MPS-65-27
27V
2.4A
0 ~ 2.7A
64.8W
100mVp-p
MPS-65-48
48V
1.35A
0 ~ 1.5A
64.8W
100mVp-p
72W(+3.3V:50W;+5V:69W)with 18CFM min. Forced air convection
6.75 ~ 8.25V 10.8 ~ 13.2V 12.2 ~ 14.85V 13.5 ~ 16.5V 21.6 ~ 26.4V 24.3 ~ 29.7V 43.2 ~ 52.8V
3.0%
2.0%
2.0%
2.0%
2.0%
2.0%
2.0%
RIPPLE & NOISE (max.)
Note.2
80mVp-p
2.97 ~ 3.63V 4.5 ~ 5.5V
3.0%
VOLTAGE TOLERANCE
Note.3
3.0%
LINE REGULATION
1.0%
1.0%
LOAD REGULATION
3.0%
3.0%
1.0%
3.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
1.0%
2.0%
SETUP, RISE, HOLD TIME
VOLTAGE RANGE
FREQUENCY RANGE
INPUT
EFFICIENCY(Typ.)
AC CURRENT
INRUSH CURRENT(max.)
LEAKAGE CURRENT
OVER LOAD
PROTECTION
800ms, 30ms, 20ms/230VAC
800ms, 30ms, 12ms/115VAC at full load
90 ~ 264VAC
47 ~ 440Hz
66%
1.6A/115VAC
127 ~ 370VDC
74%
76%
0.9A/230VAC
40A/230VAC
77%
78%
79%
80%
80%
80%
COLD START 20A/115VAC
<0.3mA / 264VAC
73 ~ 105W (3.3V:51 ~ 75W)(5V:70 ~ 105W)
rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8 ~ 4.46V
5.75 ~ 6.75V 8.63 ~ 10.1V 13.8 ~ 16.2V 15.5 ~ 18.2V 17.25 ~ 20.25V 27.6 ~ 32.4V 31 ~ 36.45V
55.2 ~ 64.8V
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
WITHSTAND VOLTAGE
Protection type : Hiccup mode, recovers automatically after fault condition is removed
-10 ~ +60 (Refer to output load derating curve)
20 ~ 90% RH non-condensing
-20 ~ +85 , 10 ~ 95% RH
0.04%/ (0 ~ 50
10 ~ 500Hz, 2G 10min./1cycle, Period for 60min.each along X, Y, Z axes
UL2601-1, TUV EN60601-1, IEC601-1 Approved
I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:0.5KVAC
SAFETY & ISOLATION RESISTANCE
I/P-O/P, I/P-FG:100M Ohms/500VDC
EMC
EMI CONDUCTION & RADIATION
Compliance to EN55011 (CISPR11) Class B
(Note 4)
HARMONIC CURRENT
EMS IMMUNITY
MTBF
Compliance to EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN60601-1-2 Medical level, criteria A
OTHERS
DIMENSION
PACKING
359.7Khrs min. MIL-HDBK-217F (25
127*76.2*42mm (L*W*H)
0.23Kg; 54pcs/14.6Kg/1.35CUFT
)
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
5. Mounting holes M1 and M2 should be grounded for EMI purposes.

MPS-65-15 Related Products

MPS-65-15 MPS-65 MPS-65-3.3 MPS-65-7.5 MPS-65-5 MPS-65-48 MPS-65-27 MPS-65-24 MPS-65-13.5 MPS-65-12
Description 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type 65W Single Output Medical Type

Technical ResourceMore

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Robot
development
community

Index Files: 280  1301  1338  2253  587  6  27  46  12  56 
Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2026 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号