MINT1022
22 Watt Single Output Series
Medical - Universal Input Power Supply
• 90-264 VAC Input
• Overvoltage and Short Circuit Protection
• Compact Size, Low Profile 2.0” x 3.5” x 0.67”
• Approved to UL, CSA and EN Standards
• CE Compliant (LVD, EMC, RoHS)
• Low Standby Power Loss (<0.5 W)
International Safety Standard Approvals
Specifications
All specifications are typical at nominal input, full load at 25°C unless otherwise stated
Output Specifications
Output Power
Natural Convection
2
2 Watts continuous,
24 W peak for 60 sec-
onds, 10% duty cycle
Not required
1.5%, see table
100 ms max.
4
% max. dev., 1.5 msec
max. recovery to 0.5%
+/-0.03%/° C typical
Standard - See chart
F
actory set with fixed
resistors to maximize
reliability
Environmental Specifications
Thermal Performance
Relative Humidity
Maximum Altitude
Vibration
Shock
Operating ambient
(see ratings chart)
Non-operating
0-50° C
50-70° C ambient
Non-condensing
Operating
/non-operating
5Hz-500Hz
Per MIL-STD-810E
0-70° C
-40 to +85° C
Full load. See table.
derate to 50%
0-95% RH
10,000 ft./40,000 ft. max.
2.5 g rms
516.4 part IV
Minimum Load
Total Regulation
Rise Time
Transient Response
(Line and Load)
At turn-on
50% step load, 0.2A/μs
Temperature Coefficient
Overvoltage Protection
Voltage Adjustability
General Specifications
Hold-up Time
Efficiency
120VAC, 60Hz
120VAC, 22W output
15ms min. at 22W output
70-80% typical depending
on output voltage
1
5 to 100kHz, varies with line
and load
U
L/EN/IEC 60601-1 CSA22.2
No.601
100 Grams (0.20 lbs)
Input Specifications
Voltage
Frequency
Input Surge Current
Universal Input
Thermistor limited
9
0-240VAC (Nominal
100-240VAC)
47-63Hz
32A max.
1
25 μA max. (200 μA
single fault)
0.45A
0.25A
2.0 A
Switching Frequency Fixed
Safety Approvals
Maximum Weight
MTBF
Overload and short
circuit protection
Medical
Earth leakage current 264V, 60Hz
Input Current
Input Fuse
120VAC
230VAC
F1
25° C, ground benign, 300,000 hours
Telcordia
Fully protected
Cycling type power limit
SL Power Electronics Corp.
•
6050 King Drive
•
Ventura, CA 93003
•
Phone:805.486.4565
•
Fax:858.712.2040
•
Email:info@slpower.com
•
www.slpower.com
MINT1022
22 Watt Single Output Series
Medical - Universal Input Power Supply
Medical
Model
MINT1022A0505I01
MINT1022A1205I01
MINT1022A2405I01
Notes
1. Other output voltages available by special order
2. Regulation from initial setpoint for 50% load change from 50% to 0% or 100%
3. Peak output rating is 24 Watts for 1 minute
4. 20MHz Bandwidth scope, differential mode. Measured with scope probe directly across output terminals of the power supply with load terminated with 0.1 μF
5. Heatsink temperatures should not be allowed to exceed 90° C
6. Installation data is online at www.slpower.com
Voltage
Output
5V
12V
24V
Output
Current
4.40A
1.83A
0.92A
Total
Regulation
+/-1.5%
+/-1.5%
+/-1.5%
Initial Setpoint
Tolerance
+/-2.5%
+/-2.5%
+/-2.5%
Ripple/
Noise
100mV
120mV
240mV
OVP
Setpoint
6.2+/-0.6V
14.1+/-1.1V
28+/-2.5V
Condor Part Number Key
M
Medical
I
Internal
N
T
1
022
Output
Wattage
A
Initial Design
of the Product
05
Output
Voltage
05
Output
Connector
I
01
Standard Factory
No. of
Product
Designation Outputs
Input
Standard
Connector Configuration
General Specifications (continued from p. 1)
E
MC Information
Conducted emissions
Radiated emissions
Line freq. harmonics
Voltage fluctuations
ESD Contact
ESD Air
Radiated immunity
Fast Transients (EFT)
Line surge immunity
Conducted immunity
Power freq. mag. field
Voltage dip immunity
EN55011
EN55011
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Level B
Level B
Complies
Complies
6 kV contact
8 kV air
3 V/m
2 kV/5 kHz
1 kV diff /2 kV
common mode
3 Vrms
3 A/m
Consult Factory
Mechanical Notes
INPUT J1 MOLEX 22-43-8040
INPUT
GROUND
OUTPUT
J1 MOLEX 22-43-8040
PIN 1 AC LINE
PIN 4 AC NEUTRAL
SOLDER HOLE ON BOARD
J2 MOLEX 22-43-8040
PIN 1 OUTPUT 1
PIN 2 OUTPUT 1
PIN 3 COMMON
PIN 4 COMMON
CONTACTS 08-70-1040
CONTACTS 08-70-1040
MATING CONNECTOR (MOLEX)
INPUT
HOUSING 50-37-5043
OUTPUT
HOUSING 50-37-5043
SL Power Electronics Corp
•
6050 King Drive
•
Ventura, CA 93003
•
Phone:805.486.4565
•
Fax:805.487.8911
•
Email:info@slpower.com
•
www.slpower.com
Data Sheet © 2008 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication.
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.
Rev.9/08/08