DTMPU63 Series
Rev. A
63 Watt
Single Output
AC/DC Medical Desktop Power Supply
FEATURES
●
Splash Proof
●
3 Year Warranty
●
Class I Insulation
●
IEC 320-C14 Input Inlet
●
Optional Output Connectors
●
CEC and Energy Star Compliant
●
Output Voltage Protection (Crowbar Design)
●
Wide Input Voltage 90 to 264VAC, 47 to 63Hz
●
Output Voltages Available from 5VDC to 36VDC
●
Input Surge Current, Over Voltage, and Over Load Protection
SPECIFICATIONS:
DTMPU63 Series
All specifications are based on 25 C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
o
SPECIFICATION
INPUT (V
in
)
Input Voltage
Input Frequency
Input Current (Low Line)
Input Current (High Line)
Inrush Current (Low Line)
Inrush Current (High Line)
Safety Ground Leakage Current
Start-Up Time
OUTPUT (V
o
)
Output Voltage Range
Load Regulation
Line Regulation
Output Power
Output Current Range
Ripple & Noise (peak to peak)
Transient Response
Hold-Up Time
Temperature Coefficient
PROTECTION
Over Voltage Protection
Over Current Protection
GENERAL
Efficiency
Dielectric Withstanding Voltage
For Primary to Secondary
Dielectric Withstanding Voltage
For Primary to Ground
Isolation Resistance
ENVIRONMENTAL
Operating Temperature
Storage Temperature
Relative Humidity
MTBF
PHYSICAL
Weight
Dimensions
Warranty
SAFETY
EMI Requirements for CISPR-11
EMI Requirements for FCC PART-18
TEST CONDITIONS
Min
90
47
Nom
Max
264
63
1.62
0.72
15
30
0.3
2
Unit
VAC
Hz
A
A
A
A
mA
s
VDC
%
%
W
A
%
ms
ms
%/°C
%
%
%
VDC
VDC
MΩ
Io = Full Load, Vin = 115VAC
Io = Full Load, Vin = 230VAC
Io = Full Load, 25°C, Cool Start, Vin = 115VAC
Io = Full Load, 25°C, Cool Start, Vin = 230VAC
Io = Full Load, Vin = 240VAC
Io = Full Load, Vin = 100VAC
0.3
12
26
0.1
1
See Rating Chart
3
0.5
Vin = 230VAC
Io = Full Load
Vin = 90 to 264VAC
Full Load, Vin = 90VAC
Io = Full Load to Half Load, Vin = 100VAC
Io = Full Load, Vin = 110VAC
All Outputs
0
See Rating Chart
0.5
16
-0.04
112
110
7
1
63
1
4
+0.04
132
150
Io = Full Load, Vin = 230VAC
Primary to Secondary
Primary to Ground
Test Voltage = 2100VDC
Derate linearly from 100% Load at 50°C to 50% load at 70°C
77
5600
2800
50
0
-40
5
85
88
50
+70
+85
95
100,000
Approximately 510 ~ 560
5.75(L) x 2.99(W) x 1.69(H)
3
Vin = 220VAC
Vin = 110VAC
B
B
°C
°C
%
hours
grams
inches
years
Class
Class
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
DTMPU63 Series
Rev. A
63 Watt
Single Output
AC/DC Medical Desktop Power Supply
OUTPUT VOLTAGE / CURRENT RATING CHART
Model Number
DTMPU63-102
DTMPU63-103
DTMPU63-104
*DTMPU63-105
*DTMPU63-106
*DTMPU63-107
*DTMPU63-108
*DTMPU63-109
*DTMPU63-110
Output Voltage
5 VDC
7 VDC
9 VDC
12 VDC
15 VDC
18 VDC
24 VDC
30 VDC
36 VDC
Max. Output Current
9A
7.85A
6.44A
5.25A
4.20A
3.50A
2.62A
2.10A
1.75A
Total Regulation
(3)
7%
7%
5%
5%
5%
5%
3%
3%
3%
Max. Output Power
45W
55W
58W
63W
63W
63W
63W
63W
63W
NOTES
1. The asterisk “
*
” next to the model number indicates CEC level IV.
2. The models with output voltages under 15VDC have been approved by TUV/PSE and models with output
voltages from 15VDC~30VDC have been approved by JET/PSE.
3. Models DTMPU63-102~103 need to use AWG#16/5C/4FT output cable in order to meet the total regulation.
Models DTMPU63-104~105 need to use AWG#16/2C/4FT output cable in order to meet the total regulation.
Models DTMPU63-106~107 need to use AWG#18/2C/4FT output cable in order to meet the total regulation.
Models DTMPU63-108~110 need to use AWG#18/2C/6FT output cable in order to meet the total regulation.
The regulation and efficiency will change if a different output cable is used.
4. Optional output connectors are available (see “DC Output Plug Selector List” link located at the bottom of the
“Desktop” category page).
DERATING CURVE
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797