Total Power International, Inc.
Tota
TPMPU31 SERIES
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single to Triple Output
Over Voltage and Over Load protection
ANSI/AAMI ES 60601-1: 2005(UL/cUL 3rd Edition)
EN 60601-1:2006 (TUV/T-mark 3rd Edition)
Input to Output : 2MOPP
Energy Star 2.0, Efficiency level V
Class I
3 year warranty
30W Desk Top Type
Medical Power Supplies
Electrical Characteristics:
Vin
fin
Po
Vo
Io
Iil
Iih
Ir
Eff
REG-i
OVP
OCP
Ttr
Th
Ts
Vp-p
Ilk
TC
OA
Pno
Vps
Vpg
Ris
Safety Approvals Input Voltage Range
Operate Voltage Range
Input Frequency
Output Power Range
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Line Regulation
Over Voltage Protection
Over Current Protection
Time of Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise(Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
Operating Altitude
No-Load Power Consumption
Dielectric Withstanding Voltage
Dielectric Withstanding Voltage
for Primary to PE
Isolation Resistance
No load, Vin=230VAC
Primary to secondary
Primary to PE
Test Voltage=500VDC
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Vin=240VAC/60Hz
All output
Io=Full load, Vin=100VAC
Io=Full load, Vin=240VAC
Io=Full load, 25°C, Cool start, Vin=115VAC
Io=Full load, 25°C, Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
Vin=230VAC
100~240VAC
90~264VAC
47~63Hz
See rating chart
See rating chart
See rating chart
0.9A
0.34A
14A (max.)
28A (max.)
67.6~85%
1% (max.)
2~10%
112~132%
110~150%
4mS (max.)
12mS (min.)
0.3~2S
1~2% (max.)
0.1mA (max.)
±0.04%/°C
Up to 3000m
See rating chart
6802VDC (min.)
2121VDC (min.)
50M
(min.)
EN 60601-1
IEC 60601-1
Application:
- Blood Pressure system
- Portable medical devices
- ECG machine
Safety Approvals:
REG-o Load Regulation
RoHS
2002/95/EC
COMPLIANT
Note:
The Ripple & Noise which is under 3.3VDC at 2% max.
The specifics for testing the energy efficiency of
TPMPU31
Series are outlined in a separate document titled
"Test Method for Calculating the Energy Efficiency of Single-Voltage External Ac-Dc and Ac-Ac Power Supplies
(August 11, 2004)," which is available on the ENERGY STAR Website.
Environmental
To
Ts
Ho
Hr
MTBF
Pd
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Temperature at 25°C
,
Calculated per MIL-HDBK-217F
Derate linearly from 100% load at 50°C to 50% load at 70°C
See derating curve
-40~85°C
0~95%
0~95%
0.1M Hrs (min.)
Total Power In
International, Inc
Toll Free: 877-646-0900
www.total-power.com
Total Power International, Inc.
Tota
TPMPU31 SERIES
Output Voltage And Current Rating Chart ( Single Output ) :
Model Number
TPMPU31-101
TPMPU31-102
TPMPU31-103
TPMPU31-104
TPMPU31-105
TPMPU31-106
TPMPU31-107
TPMPU31-108
TPMPU31-109
TPMPU31-110
30W Desk Top Type
Medical Power Supplies
Output Voltage
3 ~ 5 VDC
5 ~ 6 VDC
6 ~ 8 VDC
8 ~ 11 VDC
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 40 VDC
Output Current
6.66 ~ 4.00 A
5.00 ~ 4.16 A
4.16 ~ 3.12 A
3.75 ~ 2.72 A
2.72 ~ 2.30 A
2.30 ~ 1.87 A
1.87 ~ 1.42 A
1.42 ~ 1.11 A
1.11 ~ 0.90 A
0.90 ~ 0.75 A
Total
Regulation
7%
5%
5%
4%
3%
3%
3%
2%
2%
2%
Max
.
Output Power
20W
25W
25W
30W
30W
30W
30 W
30 W
30 W
30 W
Pno
(max.)
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
0.3W
Output Voltage And Current Rating Chart ( Multi Output ) :
Model Number
TPMPU31-200
TPMPU31-201
TPMPU31-202
TPMPU31-203
TPMPU31-204
TPMPU31-209
TPMPU31-210
TPMPU31-215
TPMPU31-301
TPMPU31-302
TPMPU31-303
TPMPU31-304
TPMPU31-305
TPMPU31-306
MPU15C-110
Output #1
Vonom Iomin Iomax Regmax
+3.3V
+5V
+5V
+5V
+3.3V
+12V
+15V
+5V
+5V
+5V
+5V
+5V
+5V
+3.3V
0.30A
0.30A
0.30A
0.30A
0.30A
0.20A
0.15A
0.30A
0.25A
0.25A
0.25A
0.30A
0.30A
0.30A
3.0A
3.0A
3.0A
3.0A
3.0A
2.0A
1.5A
3.0A
2.5A
2.5A
2.5A
3.0A
3.0A
3.0A
7%
5%
5%
5%
7%
5%
5%
5%
5%
5%
5%
5%
5%
7%
+12V
+12V
+15V
+24V
+5V
Output #2
Vonom Iomin Iomax Regmax
0.13A
0.13A
0.10A
0.07A
0.16A
1.3A
1.3A
1.0A
0.7A
1.6A
5%
5%
5%
5%
5%
-12V
-15V
-24V
-5V
-12V
-15V
-24V
-12V
-5V
Output #3
Vonom Iomin Iomax Regmax
Max. Output
Powe
r
25W
30W
30W
30W
17.9W
30W
30W
30W
25W
30W
30W
30W
30W
25W
Pno
(
max.
)
3W
3W
3W
3W
3W
3W
3W
3W
3W
3W
3W
3W
3W
3W
+12V
+12V
+15V
+24V
+24V
+12V
0.11A
0.10A
0.10A
0.10A
0.10A
0.11A
1.1A
1.0A
1.0A
1.0A
1.0A
1.1A
5%
5%
5%
5%
5%
5%
0.05A
0.05A
0.10A
0.05A
0.05A
0.05A
0.05A
0.05A
0.05A
0.5A
0.5A
1.0A
0.5A
0.5A
0.5A
0.5A
0.5A
0.5A
10%
10%
10%
10%
10%
10%
10%
10%
10%
TPMPU31-101~104
are required to use AWG#16 / 4FT output cable.
TPMPU31-105~108
are required to use AWG#18 / 6FT output cable.
TPMPU31-109~110
are required to use AWG#20 / 6FT output cable.
The regulation and efficiency will be changed by modified output cable.
Mechanical Specifications:
Note:
1. Dimensions are shown in mm.
2. Weight: 400~460gs approx.
3. Optional output connector:
See page Appendix.
Derating Curve :
120%
100%
SWITCHING
POWER
Output Watts (%)
1. Operating Temperature: 0 to 70°C
2. Derate linearly from 100% load at 50°C to 50%
load at 70°C
80%
60%
40%
20%
0%
0
10
20
30
40
50
60
70
80
Temperature (deg c)
Total Power In
International, Inc
Toll Free: 877-646-0900
www.total-power.com