LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and
workmanship for one year from the date of purchase. This
warranty does not cover fuses, disposable batteries, or
damage from accident, neglect, misuse, alteration,
contamination, or abnormal conditions of operation or handling.
Resellers are not authorized to extend any other warranty on
Fluke’s behalf. To obtain service during the warranty period,
contact your nearest Fluke authorized service center to obtain
return authorization information, then send the product to that
Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM
ANY CAUSE OR THEORY. Since some states or countries do
not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability
may not apply to you.
Fluke Corporation
Fluke Europe B.V.
P.O. Box 9090
P.O. Box 1186
Everett, WA 98206-9090
5602 BD Eindhoven
U.S.A.
The Netherlands
11/99
®
FS17XX/FS17XX IP65
AC Current Probes
Instruction Sheet
Introduction
The FS17XX/FS17XX IP65 current probe sets (the Probe) are
for accurate non-intrusive measurement of ac currents and are
used with Fluke 1735 and all 1740 products. The FS17XX IP65
can be used in more rugged environments than either the 1735
or 1745. Integrated memory for calibration data provides
current ranges from 0.44 A up to 3000 A in a frequency range
of 40 Hz to 5 kHz.
Contacting Fluke
•
•
•
•
•
•
•
Technical Support USA: 1-800-44-FLUKE (1-800-443-
5853)
Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-
5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at
www.fluke.com.
To register your product, visit
http://register.fluke.com.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Safety Instructions
Please read this section carefully. It will familiarize you with the
most important safety instructions for handling the Probe. In
this instruction sheet, a Warning identifies conditions and
actions that pose hazard(s) to the user. A Caution identifies
conditions and actions that may damage the calibrator or the
test instruments.
August 2009
©2009
Fluke Corporation, All rights reserved. Specifications are subject to
change without notice. All product names are trademarks of their respective
companies.
WX
Warning
The Probe may only be used and handled by
qualified personnel. To avoid electric shock or
personal injury, follow these precautions:
•
Do not apply to or remove from hazardous
live conductors without taking additional
protective measures.
•
High voltages and currents may be present in
adjacent circuits under test.
•
Do not use the Probe if damaged. Always
connect to display device before it is installed
around the conductor.
•
Use the Probe only as specified in the
operating instructions; otherwise the safety
features may not protect you.
•
Adhere to local and national safety codes.
Individual protective equipment must be used
to prevent the shock and arc blast injury
where hazardous live conductors are
exposed.
•
Before each use, inspect the Probe. Look for
cracks or missing portions of the housing or
output cable insulation. Also look for loose or
weakened components.
•
Use caution when working with voltages
above 60 V dc, 30 V ac rms or 42 V ac peak.
Such voltages pose a shock hazard.
•
Equipment is considered to be used in 600 V
CAT IV and 1000 V CAT III environment.
•
CAT IV equipment is designed to be used at
the source of distribution system up to the
energy-meter within the installation. CAT III
equipment is designed to be used in
distribution panels, feeders and short branch
circuits, and the lighting system in large
buildings.
•
Do not use Probe in damp or wet
environments or in locations that hazardous
gases exist.
Operating Instructions
1.
2.
Connect the Probe to the product current input jack.
Wear protective gloves or de-energize the circuit and
place the Probe around the conductor under test. Re-
energize the circuit.
Observe and take measurements as required.
Positive output indicates that the current flow is in the
direction shown by the arrow on the Probe.
Wear protective gloves or de-energize the circuit
before removal of the Probe.
Electrical Characteristics
Input ranges I
I
L1, L2, L3, N:
15 /150 / 1500 / 3000 A AC
Measuring range:
0.44 A to 3000 A AC
Intrinsic error:
<2 % of I
I
Position influence:
max.
±2
% of m.v. for distance conductor
to measuring head >30 mm (1.18 in)
Stray field influence:
<±2 A for Iext= 500 A AC and distance
to measuring head >200 mm (7.87 in)
Temperature coefficient:
<0.05 % / K
Current transformer:
ratio :
≤999
kA /
≤I
I
Ratio selection:
by job programming for 174x series; by setup
menu for 1735
Connection:
3-phase, 3-phase + N; 2-phase L1 and L3 (2W-
meter-method); 7-pole connector
Specifications
General Specifications
Cable length:
2 m (79 in)
Length of measuring head:
61 cm (24 in)
Operating temperature range:
-10
°C
to +70
°C
(14
°F
to 158
°F)
Protection Class:
FS17XX IP65-IP65, FS17XX- IP41
Operating Humidity:
10 % to 80 %, non- condensing
Storage temperature range:
-20
°C
to +90
°C
(-4
°F
to 198
°F)
Weight:
.92 kg (2.03 lb)
EMC Standards:
IEC/EN 61326-1:2006
Reference Conditions
Environment temperature range:
+18
°C
to +26
°C
(64.4
°F
to 78.8
°F)
Humidity:
20 to 75 % rh
Altitude:
max. 2000 m (1.24 mi) for 1000 V CAT III / 600 V
CAT IV
max. 5000 m (3.107 mi) for 600 V CATIII / 300 V CATIV
Current:
upper value of selected range sinusoidal waveform,
48 Hz to 65 Hz, distortion factor: <1 %, no DC component,
stray field <40 A/m, conductor centered within the Probe.
3.
4.
Safety Standards
IEC/EN61010-1:2001
IEC/EN61010-2-032:2002
IEC/EN61010-031:2002
Cleaning
Clean the Probe periodically by wiping it with a damp cloth and
detergent. Do not use abrasive cleaners or solvents. Do not
immerse the Probe in liquids.
Symbols
Safety Specifications
Category Rating:
1000 V CAT III, 600 V CAT IV, pollution
degree 2. The application of the Probe on
uninsulated
conductors
is limited to 1000 V ac rms or dc to ground and
frequencies below 1 kHz.
~
-
T
W
)
P
Do not dispose of this product as unsorted municipal
waste. Go to Fluke’s website for recycling information.
Do not apply around or remove from HAZARDOUS LIVE
conductors, without using protective gloves.
Product is protected by double insulation.
Risk of Danger. Important information. See Instruction
Sheet.
Conforms to relevant Canadian Standards Association
directives.
Conforms to relevant European Union directives.