• over voltage, overload and short circuit protections
• medical 60601-1 4th edition safety approvals
• designed for 2 x MOPP applications
• active PFC
• level V efficiency
• custom designs available
RoHS
MODEL
ETMA1201666U
ETMA1601250U
ETMA1901053U
ETMA2001000U
ETMA240833U
ETMA360555U
ETMA480416U
Notes:
output
voltage
(Vdc)
output
current
max
(A)
output
power
max
(W)
ripple
1
max
(mVp-p)
efficiency
level
12
16
19
20
24
36
48
16.66
12.5
10.53
10
8.33
5.55
4.16
200
200
200
200
200
200
200
240
300
300
300
300
300
300
V
V
V
V
V
V
V
1. At full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with 10 µF aluminum electrolytic and 0.1 µF ceramic capacitors.
PART NUMBER KEY
ETMA1201666U
C
-
P51
-
WP
-
CXX
Base Number
Reserved for Custom
Configurations
Factory
Designation
example of 12 Vdc, 16.66 A
DC Plug Type
Input Cable:
C = included
blank = no cable
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 200W U
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
06/20/2017
│
page
2 of 5
INPUT
parameter
voltage
frequency
current
inrush current
no load power consumption
at 90 Vac, 60 Hz
at 230 Vac, cold start
conditions/description
min
90
47
typ
max
264
63
3
150
0.5
units
Vac
Hz
A
A
W
OUTPUT
parameter
line regulation
load regulation
hold-up time
temperature coefficient
at 115 Vac
16
±0.05
conditions/description
min
typ
±1
±5
max
units
%
%
ms
%/°C
PROTECTIONS
parameter
conditions/description
12 V output
16, 20 V outputs
19 V output
24, 36 V outputs
48 V output
shutdown and auto restart
continuous, auto restart
110
110
111
110
110
110
133
130
137
125
120
160
%
%
%
%
%
%
over voltage protection
overload protection
short circuit protection
SAFETY & COMPLIANCE
parameter
isolation voltage
insulation resistance
safety approvals
EMI/EMC
leakage current
MTBF
RoHS compliant
conditions/description
input to output for 2 seconds
input to frame ground for 2 seconds
input to output, at 500 Vdc
input to frame ground, at 500 Vdc
20
20
min
typ
max
5,656
2,121
units
Vdc
Vdc
MΩ
MΩ
medical UL/cUL (UL 60601-1), EN 60601-1, IEC 60601-1 4th edition
FCC 47 CFR Part 18, CE, EN 61000-3-(2, 3), EN 55011, IEC 61000-4-(2, 3, 4, 5, 6, 8, 11)
earth, at 264 Vac
enclosure, at 264 Vac
at 25°C, max. load
yes
140,000
190
10
300
100
µA
µA
hours
ENVIRONMENTAL
parameter
operating temperature
storage temperature
humidity
non-condensing
conditions/description
min
0
-10
20
typ
max
50
70
90
units
°C
°C
%
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 200W U
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
06/20/2017
│
page
3 of 5
MECHANICAL
parameter
dimensions
input plug
weight
conditions/description
7.42 x 4.11 x 2.36 (188.5 x 104.5 x 60 mm)
IEC320 / C14
1.06
kg
min
typ
max
units
inch
MECHANICAL DRAWING
units: mm
104.5 0.5
60.0 0.5
188.5 0.5
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 200W U
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
06/20/2017
│
page
4 of 5
DC CORD
PIN 2
Blue (GND)
PIN 1
Red (+Vo)
[2464#16AWG 4C+1(16AWG)]
PIN 4
Black (GND)
Core
14.2*8.0*28.5mm
SECTIONAL DRAWING
D.W. (GND)
PIN 3
Yellow (+Vo)
FRONT VIEW
4P POWER DIN
TERMINAL
ø3.2*2
Black+Blue+D.W.
Yellow+Red
UL Heat Shrinkable Tube
4PIN POWER DIN
COMPONENT LIST
1. Cable: color black, UL/CSA 2464 80/90°C 300V
16 AWG X 4C+Aluminum Shield (16AWG)
2. PVC Wire: color: blue/black/yellow/red
3. Connector: 4PIN power DIN or EQ
4. 4PIN power DIN (female) MDP-002-4PIN (P530600004F0001)
AC CORD
cui.com
For more information, please visit the
product page
|
For information on related products, view
Dc Power Jacks
CUI Inc
│
SERIES:
ETMA 200W U
│
DESCRIPTION:
MEDICAL AC-DC POWER SUPPLY
date
06/20/2017
│
page
5 of 5
REVISION HISTORY
rev.
1.0
1.01
1.02
1.03
description
initial release
V-Infinity branding removed, safety and EMI/EMC data updated
replaced TUV approval with Nemko
updated to medical 60601-1 4th edition
date
03/05/2012
08/13/2012
09/04/2013
06/20/2017
The revision history provided is for informational purposes only and is believed to be accurate.
Headquarters
20050 SW 112th Ave.
Tualatin, OR 97062
800.275.4899
Fax 503.612.2383
cui.com
techsupport@cui.com
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause undesired operation.
CUI offers a two (2) year limited warranty. Complete warranty information is listed on our website.
CUI reserves the right to make changes to the product at any time without notice. Information provided by CUI is believed to be accurate and reliable. However, no responsibility is
assumed by CUI for its use, nor for any infringements of patents or other rights of third parties which may result from its use.
CUI products are not authorized or warranted for use as critical components in equipment that requires an extremely high level of reliability. A critical
component is any component of a life support device or system whose failure to perform can be reasonably expected to cause the failure of the life support device or system, or to