40W AC-DC
Reliable Green
Medical Adaptor
GSM40B
series
K60950-1
(Only GSM40B48-P1J)
J60950-1
(Only GSM40B48-P1J)
IEC60950-1
+ +
(
for 5~9V)
Features
‧
Ⅱ
‧
‧
‧
‧
‧
‧
℃
‧
‧
‧
‧
‧
‧
‧
‧
‧
Description
μ
{
ψψ
File Name:GSM40B-SPEC 2018-07-18
40W AC-DC
Reliable Green
Medical Adaptor
SPECIFICATION
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER (max.)
OUTPUT
GSM40B
GSM40B15
15V
2.67A
0 ~ 2.67A
40W
100mVp-p
GSM40B18
18V
2.22A
0 ~ 2.22A
40W
120mVp-p
GSM40B24
24V
1.67A
0 ~ 1.67A
40W
150mVp-p
series
GSM40B05-P1J GSM40B07-P1J GSM40B09-P1J GSM40B12-P1J GSM40B15-P1J GSM40B18-P1J GSM40B24-P1J GSM40B48-P1J
GSM40B05
Note.2
5V
5A
0 ~ 5A
25W
GSM40B07
7.5V
5.34A
0 ~ 5.34A
40W
80mVp-p
GSM40B09
9V
4.45A
0 ~ 4.45A
40W
100mVp-p
GSM40B12
12V
3.34A
0 ~ 3.34A
40W
100mVp-p
GSM40B48
48V
0.84A
0 ~ 0.84A
40W
150mVp-p
RIPPLE & NOISE (max.)
Note.3
80mVp-p
VOLTAGE TOLERANCE
Note.4
±5.0%
LINE REGULATION
Note.5
±1.0%
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE
FREQUENCY RANGE
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
OVERLOAD
±5.0%
±5.0%
±1.0%
±5.0%
±5.0%
±1.0%
±5.0%
±3.0%
±1.0%
±3.0%
±3.0%
±1.0%
±3.0%
±3.0%
±1.0%
±3.0%
±2.5%
±1.0%
±2.5%
±2.5%
±1.0%
±2.5%
Note.6
1000ms, 30ms / 230VAC
1500ms, 30ms / 115VAC at full load
50ms / 230VAC
24ms / 115VAC at full load
Note.7
80 ~ 264VAC 113 ~ 370VDC
47 ~ 63Hz
81%
1A / 115VAC
85.5%
0.5A / 230VAC
86%
88%
88.5%
89%
90%
91%
INPUT
Cold start
30A / 115VAC
60A / 230VAC
Touch current < 50
μ
A/264VAC
105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.2 ~ 7,0V
7.8 ~ 10.2V
9.4 ~ 12.2V
12.6 ~ 16.2V 15.7 ~ 20.3V
18.9 ~ 24.3V
Protection type : Shut down o/p voltage, re-power on to recover
-30 ~ +70
℃
(Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85
℃
, 10 ~ 95% RH non-condensing
PROTECTION
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
25.2 ~ 32.4V
50.4 ~ 64.8V
±
0.03% /
℃
(0 ~ 50
℃
)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
OPERATING ALTITUDE
Note.8
3000 meters
IEC60601-1, EN60601-1
/
EN
60601-1-11
, ANSI/AAMI ES60601-1
/ ES60601-1-11
(3.1 version),PSE
J60950-1,KC K60950-1,
SAFETY STANDARDS
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3
,
EAC TP TC 004
approved
ISOLATION LEVEL
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
Primary-Secondary: 2xMOPP
I/P-O/P:4KVAC
I/P-O/P:100M Ohms / 500VDC / 25
℃
/ 70% RH
Parameter
Standard
EN55011 (CISPR11), FCC PART 15 / CISPR22,
Conducted emission
CAN ICES-3(B)/NMB-3(B)
,MSIP KN32
EN55011 (CISPR11), FCC PART 15 / CISPR22,
Radiated emission
CAN ICES-3(B)/NMB-3(B)
,MSIP KN32
EN61000-3-2
Harmonic current
Voltage flicker
EN55024 , EN60601-1-2, EN61204-3
Parameter
ESD
RF field susceptibility
EMC IMMUNITY
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
MTBF
DIMENSION
PACKING
PLUG
CABLE
740K hrs min. MIL-HDBK-217F(25
℃
)
125*50*31.5mm (L*W*H)
0.29Kg; 40pcs/12.6Kg/1.05CUFT
See page 4~5 ; Other type available by customer requested
See page 4~5 ; Other type available by customer requested
EN61000-3-3
Standard
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Class B
Class B
Class A
-----
Test Level / Note
Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
Level 3, 2KV
Level 3, 1KV/Line-Line
Level 3, 10V
Level 4, 30A/m
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
EMC EMISSION
SAFETY &
EMC
(Note 9)
OTHERS
CONNECTOR
NOTE
℃
μ
μ
℃℃
For
guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM40B-SPEC 2018-07-18