V10
TR100M SERIES
100W MEDICAL SWITCHING
ADAPTER
Features
* Universal Input Range 90~264VAC
* Meets EN60601-1 and EN55011 Class B
* Continuous Short Circuit Protection
* Over Voltage Protection
* Efficiency & Standby Power Meet Level V
(TR100M120~150: Output Cable Length
≦
1500mm 14AWG /UL1185)
(TR100M180~480: Output Cable Length
≦
1800mm 16AWG /UL1185)
* Meets 2 MOPP
Ordering information
TR100MXXX- XX
Model No.
DC Plug Type
X
OVP
E: with OVP
XX or XXX
DC Cable Length and Type
11: 720mm with Ferrite Core
12: 1220mm with Ferrite Core
36: 1500mm with Ferrite Core
13: 1800mm with Ferrite Core
MODEL
OUTPUT
VOLTAGE
12 V
15 V
18 V
19 V
20 V
24 V
48 V
OUTPUT
CURRENT
8.34 A
6.67 A
5.56 A
5.27 A
5.0 A
4.17 A
2.1 A
RIPPLE &
NOISE
NOTE 2
VOLTAGE
ACCURACY
NOTE 1
LINE
REGULATION
NOTE 3
LOAD
REGULATION
NOTE 4
% EFF.
(Typ.)
NOTE 5
TR100M120
TR100M150
TR100M180
TR100M190
TR100M200
TR100M240
TR100M480
1%
1%
1%
1%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±4%
±3%
±2%
±2%
±2%
±2%
±2%
88%
88%
88%
88%
88%
89%
89%
1%
1%
1%
Specifications are subject to change without notice.
1
V10
Specifications
INPUT SPECIFICATIONS:
Voltage …....………....…..............…………………..……………… 90~264Vac
Frequency ………………………………..………………………..... 47 to 63Hz
Inrush Current ……………………… Cold Start @25℃ 100A max. @240Vac
Conducted EMI …………………………..……………… CISPR/FCC Class B
Isolation ……………………………………………… Input to output = 4k Vac
Leakage Current …………………………………………………… 0.1mA max.
TR100M Series Derating Curve
OUTPUT SPECIFICATIONS:
Hold-up Time ………………………………………......… 16ms typ. @115Vac
Short Circuit Protection …………….....……….. Hiccup Mode (Auto Recover)
Over Voltage Protection …………….…...……..... TVS Component to Clamp
Temperature Coefficient …………………….................................. ±0.05%/℃
GENERAL SPECIFICATIONS:
Operating Temperature ............................... -20 ~ 60℃ (see derating curve)
Storage Temperature …………………………...….......……........-20 ~ 85℃
Humidity …………………..….......………........ 93% RH max. Non condensing
Cooling ………………………………….…………………… Natural Convection
Switching Frequency ……………………….…………………… 70KHz Typical
MTBF ……… MIL-HDBK-217F, GB, at 25℃/115VAC ………. 150Khrs min.
Altitude ………………………………………….………………………… 3000m
Dimensions …….……. 5.591x2.283x1.457 inches (142.00x58.00x37.00mm)
Weight …………………... ……………………………………………………485g
AC Inlet ………………………………. …………………… …………IEC320/C8
SAFETY AND EMC:
Emission and lmmunity ………… EN55011, EN60601-1-2, EN61000-3-2
EN61000-3-3
Safety (Medical 3 )……………………………………… IEC60601-1, EN60601-1
UL ANSI/AAMI ES60601-1
rd.
NOTE:
1. Voltage setpoint at 60% full load.
2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for
Ripple & Noise measuring @20MHz BW.
3. Line regulation measured from 100Vac to 240Vac, full load.
4. Load regulation measured from 60% to 100% full load and from 60%
to 20% load (60% +/- 40% full load)
5. Typical efficiency with 230 VAC and max. load at 25°
C.
Mechanical Specification
Typical at 25˚C, nominal line and 75% load, unless otherwise Specified
2