EEWORLDEEWORLDEEWORLD

Part Number

Search

GNT30-28

Description
Switching Power Supplies 30W 28V@1.07A
CategoryPower/power management   
File Size305KB,2 Pages
ManufacturerSL Power
Download Datasheet Parametric Compare View All

GNT30-28 Online Shopping

Suppliers Part Number Price MOQ In stock  
GNT30-28 - - View Buy Now

GNT30-28 Overview

Switching Power Supplies 30W 28V@1.07A

GNT30-28 Parametric

Parameter NameAttribute value
Product AttributeAttribute Value
ManufacturerSL Power
Product CategorySwitching Power Supplies
RoHSN
Output Voltage-Channel 128 VDC
Number of Outputs1 Output
Output Power30 W
Open Frame/EnclosedOpen Frame
Input Voltage90 VAC to 264 VAC
Output Current-Channel 1800 mA
IndustryMedical
Mounting StyleChassis
Length3.15 in
Width1.95 in
Height0.98 in
ApprovalsEN60601-1
Operating Temperature Range0 C to + 50 C
Cooling MethodConvection, Air-Cooled
Input Voltage, Max264 VAC
Input Voltage, Min90 VAC
Input Voltage, Nominal100 VAC to 240 VAC
Load RegulationRegulated
Mechanical StyleOpen / Enclosed Frame
Output Voltage V1, Nominal28 VDC
Factory Pack Quantity30
Unit Weight3.520000 oz
GNT30 MEDICAL SERIES
Low Power Medical Power Supplies 30 Watt AC/DC Universal Input Power Supply
SINGLE OUTPUT POWER SUPPLY
SUMMARY
• 90-264 Vac input range
• Overvoltage and short circuit protection
• Approved to UL, CSA and EN Standards
• Small 50 mm x 80 mm footprint
• Class II input available (G2T30)
The GNT30 Series is a 30 W universal input AC/DC Power
supply in a very small footprint. With the medical approvals
the GNT30 is ideal for a variety of medical device applications
including, small single board computers, battery charging, and
running small motors, pumps, and solenoids. The series with full
approval to EN60601-1 Standard improves design-in time, and
reduces end system compliance costs.
All specifications are typical at nominal input, full load at 25
O
C unless otherwise stated
SPECIFICATIONS
OUPUT SPECIFICATIONS
Output Power
Total Regulation
Turn-on time
Transient response
Temperature Coef-
ficient
Overvoltage protection
Short Circuit protection
Power cycling
0.25 A/ms 50% load
deviation
150 l.f.m. airflow
30 Watts max.(See note 4
See table
1.5 s max.
3.5% max. dev. 1 ms recovery
to 1%
+/-0.03%/oC
125% +/-10%
Yes
INPUT SPECIFICATIONS
Input Voltage range
Input frequency
Input surge current
Ground Leakage current
Input current
Input fuse
132 Vac 60 Hz
120 Vac
230 Vac
F1,F2
Universal input
100-240 nom (90 - 264 max)
Vac
55 Hz +/- 10 Hz
32 A max.
100
mA
(180
mA
single fault)
0.6 A
0.37 A
2.5 A
EMC INFORMATION
Conducted emissions
Radiated emissions
Line freq. harmonics
Voltage fluctuations
ESD Air
ESD Contact
Radiated immunity
Fast Transients (EFT)
Line surge immunity
Conducted immunity
Power freq. mag. field
Voltage dip immunity
EN55011
EN55011
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Level B
Level A
Complies
Complies
8 kV contact
6 kV air
3 V/m
2 kV
1 kV diff /2 kV cm
3 V
rms
3 A/m
5 cycles 40 % vnom 0.5 Watts.
25 cycles 70 % vnom 20 W
ENVIRONMENTAL SPECIFICATIONS
Thermal Performance
operating ambient
(see chart)
non-operating
0 - 50
o
C
150 l.f.m (See note 4.)
50 -70
o
C ambient
150 l.f.m
Relative Humidity
Maximum Altitude
Vibration
Shock
non-condensing
operating / non-operating
Operating
Non-operating
Operating
Non-operating
0-50
o
C
-40 to +85
o
C
30 Watts
derate 2% /
o
C
0% - 95% RH
10,000 ft. / 40,000 ft. max.
1.5 g
rms
20-2000 Hz
5 g
rms
20-2000 Hz
20 g
pk
10 ms, half-sine
40 g
pk
10 ms half-sine

GNT30-28 Related Products

GNT30-28 GNT30-15 GNT30-15-G
Description Switching Power Supplies 30W 28V@1.07A Switching Power Supplies 30W 15V@2.0A Switching Power Supplies 30W 15V@2.0A
Product Attribute Attribute Value Attribute Value Attribute Value
Manufacturer SL Power SL Power SL Power
Product Category Switching Power Supplies Switching Power Supplies Switching Power Supplies
RoHS N N Details
Output Voltage-Channel 1 28 VDC 15 VDC 15 VDC
Number of Outputs 1 Output 1 Output 1 Output
Output Power 30 W 30 W 30 W
Open Frame/Enclosed Open Frame Open Frame Open Frame
Input Voltage 90 VAC to 264 VAC 90 VAC to 264 VAC 90 VAC to 264 VAC
Output Current-Channel 1 800 mA 1.06 A 2 A
Industry Medical Medical Medical
Mounting Style Chassis Chassis Chassis
Length 3.15 in 3.15 in 3.15 in
Width 1.95 in 1.95 in 1.97 in
Height 0.98 in 0.98 in 0.98 in
Approvals EN60601-1 EN60601-1 EN60601-1
Load Regulation Regulated Regulated Regulated
Operating Temperature Range 0 C to + 50 C 0 C to + 50 C -
Cooling Method Convection, Air-Cooled Convection, Air-Cooled -
Input Voltage, Max 264 VAC 264 VAC -
Input Voltage, Min 90 VAC 90 VAC -
Input Voltage, Nominal 100 VAC to 240 VAC 100 VAC to 240 VAC -
Mechanical Style Open / Enclosed Frame Open / Enclosed Frame -
Output Voltage V1, Nominal 28 VDC 15 VDC -
Factory Pack Quantity 30 30 -
Unit Weight 3.520000 oz 3.520000 oz -

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Robot
development
community

Index Files: 2876  2491  2171  1773  2709  58  51  44  36  55 
Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2026 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号