160W AC-DC Reliab le Green Medical Adaptor
GSM160A
series
EN 60601-1
IEC 60601-1
8
EN60601-1 ANSI/AAMI ES60601-1 IEC60601-1
+ +
Applications
Mobile clinical workstation
Oral irrigator
Portable hemodialysis machine
Breath Machine
Medical computer monitor
Features
3 pole AC inlet IEC320-C14, Class
Ⅰ
power unit
Medical safety approved (2 x MOPP) accroding to
ANSI/AAMI ES60601-1 and IEC/EN60601-1
Extremely low leakage current
No load power consumption<0.15W
Energy efficiency level VI and meet CoC Version 5
-30~+70
℃
wide range working temperature
Protections: Short circuit / Overload / Over voltage / Over temperature
LED indicator for power on
Lifetime > 85 K hours
3 years warranty
Description
GSM160A is a highly reliable, 160W desktop style single-output green medical adaptor series. This product
is equipped with a 3-pin (
with
FG) standard IEC320-C14 power plug, adopting the input range from 80VAC to
264VAC. The entire series supplies different output voltages between 12VDC and 48VDC that can satisfy
the demands for various kinds of medical electrical devices. The circuitry design meets the international medical
standards (2*MOPP), having an ultra low leakage current (<100 A), fitting the medical devices in direct electrical
μ
contact with the patients.
With the efficiency up to 94% and the extremely low no-load power consumption below 0.15W,
GSM160A is
compliant with USA EISA 2007/DoE, Canada NRCan, Australia and New Zealand MEPS, EU ErP,and meet
Code of Conduct (CoC) Version 5. The supreme feature allows the adaptor to
save the energy when it is either
under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic
case. GSM160A is approved with the international medical safety certificates.
Model Encoding
GSM160 A 12 - R7B
DC plug type
Output voltage
IEC320-C14 AC inlet
Rated wattage
Series name
{
R7B:
Standard model,
Power DIN 4Pin with lock type
Other options available by customer requested (see Page 4~5)
File Name:GSM160A-SPEC 2018-01-12
160W AC-DC Reliab le Green Medical Adaptor
SPECIFICATION
ORDER NO.
SAFETY MODEL NO.
DC VOLTAGE
RATED CURRENT
CURRENT RANGE
RATED POWER (max.)
OUTPUT
GSM160A12-R7B
GSM160A12
Note.2
12V
GSM160A
GSM160A24-R7B
GSM160A24
24V
6.67A
0 ~ 6.67A
160W
120mVp-p
±3.0%
±1.0%
±3.0%
48V
3.34A
0 ~ 3.34A
160W
150mVp-p
±3.0%
±1.0%
±3.0%
series
GSM160A15-R7B
GSM160A15
15V
9.6A
0 ~ 9.6A
144W
100mVp-p
±5.0%
±1.0%
±5.0%
24ms / 115VAC at full load
GSM160A20-R7B
GSM160A20
20V
8A
0 ~ 8A
160W
100mVp-p
±4.0%
±1.0%
±4.0%
GSM160A48-R7B
GSM160A48
11.5A
0 ~ 11.5A
138W
RIPPLE & NOISE (max.)
Note.3
80mVp-p
VOLTAGE TOLERANCE
Note.4
±5.0%
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE
FREQUENCY RANGE
POWER FACTOR (Typ.)
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT(max.)
OVERLOAD
Note.5
±1.0%
±5.0%
Note.6
2000ms, 50ms / 230VAC
2500ms, 50ms / 115VAC at full load
24ms / 230VAC
Note.7
80 ~ 264VAC
47 ~ 63Hz
INPUT
12V,15V:PF>0.93 / 230VAC
20V,24V,48V:PF>0.94 / 230VAC
90%
91%
92.5%
1.85A / 115VAC
1A / 230VAC
PF>0.98 / 115VAC at full load
93%
94%
Cold start 55A / 115VAC
110A / 230VAC
Earth leakage current < 115
μ
A/264VAC , Touch current <90
μ
A/264VAC
105 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
105 ~ 135% rated output voltage
Protection type : Shut down o/p voltage, re-power on to recover
Shut down o/p voltage, re-power on to recover
-30 ~ +70
℃
(Refer to "Derating Curve")
20% ~ 90% RH non-condensing
-40 ~ +85
℃
, 10 ~ 95% RH non-condensing
±
0.03% /
℃
(0~40
℃
)
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
PROTECTION
OVER VOLTAGE
OVER TEMPERATURE
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
OPERATING ALTITUDE
Note.8
3000 meters
IEC60601-1, TUV EN60601-1, ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3
,
EAC TP TC 004
approved
SAFETY STANDARDS
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP
ISOLATION LEVEL
WITHSTAND VOLTAGE
Note.9
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25
℃
/ 70% RH
Parameter
Standard
Test Level / Note
Conducted
emission
EMC EMISSION
EN55011 (CISPR11),
FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
EN55011 (CISPR11),
FCC PART 15 / CISPR22,
CAN ICES-3(B)/NMB-3(B)
EN61000-3-2
Class B
Class B
Class A
-----
Test Level / Note
Radiated emission
Harmonic current
SAFETY &
EMC
(Note 10)
EN61000-3-3
Voltage flicker
EN55024 , EN60601-1-2, EN61204-3
Parameter
ESD
Standard
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
RF field susceptibility
EMC IMMUNITY
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
Level 3, 2KV
Level 3, 1KV/Line-Line , 2KV/Line-FG
Level 3, 10V
Level 4, 30A/m
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
OTHERS
MTBF
DIMENSION
PACKING
PLUG
CABLE
CONNECTOR
NOTE
239.4K hrs min. MIL-HDBK-217F(25
℃
)
175*72*35mm (L*W*H)
0.66Kg; 20pcs/14.2Kg/1.06CUFT
See page 4~5 ; Other type available by customer requested
See page 4~5 ; Other type available by customer requested
1. All parameters are specified at 230VAC input, rated load, 25
℃
70% RH ambient.
2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1 f & 47 f capacitor.
μ
μ
4. Tolerance: includes set up tolerance, line regulation, load regulation.
5. Line regulation is measured from low line to high line at rated load.
6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
7.
Derating may be needed under low input voltages. Pleas check the derating curve for more details.
8. The ambient temperature derating of 3.5
℃
/1000m with fanless models and of 5
℃
/1000m with fan models for operating altitude higher than 2000m(6500ft).
9. Optional for 1.5KVAC with BF rated.
10. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the
EMC directives.
For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name:GSM160A-SPEC 2018-01-12