DTM110-C8 Series
110W Medical/ITE Class II External Power Supplies
Features
Meets DoE Level VI & EU Tier 2 Efficiency
Medical & ITE Certifications
Class II Input (No Earth Ground Required)
< 0.15W Off-load Power Draw
Meets IEC60601-1-2 Ed4 & 60601-1-11
Suitable for B & BF Rated Equipment
Key Market Segments & Applications
Specifications
Input Voltage range
Inrush Current
Input Current
Touch Current
Power Factor
Hold Up Time (Typ)
Temperature Coefficient
Adjustment Range
Minimum Load
Total Regulation
Ripple & Noise (pk-pk)
Overload & Short Circuit Protection
Overvoltage Protection
Efficiency
Operating Temperature
Storage Temperature
Humidity (non condensing)
Cooling
Withstand Voltage
Safety Agency Certification
Offload Power Consumption
EMC
IP Rating
Altitude
Weight (Typ)
Size (WxLxH)
Cable Length
AC Input Connector
Output Connector
Warranty
Model
VAC
A
A
-
-
ms
%/°C
-
A
%
mV
-
V
%
°C
°C
%RH
-
VAC
-
W
-
-
m
g
in (mm)
mm
-
-
yrs
DTM110-C8
90 - 264VAC (47 - 63Hz)
<120A at 230VAC input, 25°C ambient cold start
2A @90VAC; 1.25A @264VAC Maximum
<100uA @ 264VAC
Meets EN61000-3-2, >0.9 at 230VAC full load
10ms at 115VAC input
±0.05%/°C
None
None
12-19V: ±5%, 24-48V: ±3%
150mV
105 - 150% - hiccup mode
Cycle input power to reset
>89%, average efficiency
-20 to +60°C derate linearly to 50% load from 40 to 60°C
Derate linearly to 80% load from 0 to -20°C
-20 to +85°C
0 - 95%RH
Convection
Input to Output 4kVAC (2xMoPP)
IEC/ES/CSA/EN60601-1, IEC60601-1-11, IEC/UL/CSA/EN60950-1, CE Mark
< 0.15W
IEC60601-1-2 Ed4:2014, EN55011 Class B, EN55032 Class B,
FCC Part 15 Class B, FCC Part 18 Class B
IP41
5,000m maximum
780g
2.48 x 5.5 x 1.29” (63 x 140 x 33mm)
1000mm
IEC 320-C8 (2 prong)
Kycon KPPX-4P or equivalent; Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Three Years
Notes:
Other connectors are available. Contact factory.
1-800-526-2324 • www.us.tdk-lambda.com/lp/
150
Disclaimer: us.tdk-lambda.com/lp/legal.htm