EEWORLDEEWORLDEEWORLD

Part Number

Search

SDU

Description
Medical Approvals
File Size192KB,2 Pages
ManufacturerXP Power
Websitehttps://www.xppower.com/
Download Datasheet Compare View All

SDU Overview

Medical Approvals

AC DC
150 Watts
SDU Series
Medical Approvals
3.2” x 5” Footprint
Fits 1U Applications
High Efficiency
12 V & 24 V Outputs
xppower.com
U Channel or Open Frame Versions
3 Year Warranty
Specification
Input
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
No Load Input Power
Earth Leakage Current
Input Protection
90-264 VAC
47-63 Hz
2.0 A at 90 VAC, 1 A at 230 VAC
Max 100 A at 230 VAC, cold start 25 °C
>0.9 at >40% load
<1 W
<250 µA at 264 VAC
F5 A / 350 V internal in line fuse
General
Efficiency
Isolation
• Typically 87%
• 4000 VAC Input to Output
1500 VAC Input to Ground
1000 VDC Output to Ground
• Primary to secondary: 2 x MOPP
Primary to earth: 1 x MOPP
• 65 kHz PFC typical,
55 kHz - 65 kHz PWM typical at full load
• 8.4 W/in
3
• 110 kHrs to MIL-HDBK-217F at 25 °C, GB
Protection Level
Switching Frequency
Power Density
MTBF
Output
Output Voltage
Output Voltage Trim
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
Line Regulation
Load Regulation
Over/Undershoot
Transient Response
Ripple & Noise
Overvoltage Protection
Overload Protection
Short Circuit Protection
Temperature
Coefficient
See table
±10%
±1%
No minimum load required
3.0 s max at 120 VAC
30 ms typical
20 ms min at full load, 120 VAC
±0.5%
±2%
3% max
4% max deviation, recovery to within
1% in 500 µs for a 50% load change
1% pk-pk, 20 MHz bandwidth
112-130% Vnom with auto recovery
110-185%
Trip & restart (hiccup mode), auto recovery
0.04% /°C
Environmental
Operating Temperature
• 0 °C to +70 °C, derate at 2.5%/ °C from
+50 °C to +70 °C
Storage Temperature
• -20 °C to +85 °C
Operating Humidity
• 5-90%, non-condensing
Storage Humidity
• 5-95%, non-condensing
Cooling
• Convection-cooled
Operating Altitude
• 3000 m
Vibration
• 5-50 Hz, acceleration 7.35 m/s
2
on
X, Y and Z axis
EMC & Safety
Emissions
Harmonic Currents
Voltage Flicker
ESD Immunity
Radiated Immunity
EFT/Burst
Surge
Conducted Immunity
Dips & Interruptions
• EN55011 Level B conducted & radiated
• EN61000-3-2 class A
EN61000-3-2 class C for loads ≥40%
• EN61000-4-2, level 3 Perf Criteria A
• EN61000-3-3
• EN61000-4-3, 3 V/m Perf Criteria A
• EN61000-4-4, level 2 Perf Criteria A
• EN61000-4-5, installation class 3,
Perf Criteria A
• EN61000-4-6, 3V Perf Criteria A
• EN61000-4-11, 30% 500 ms, 60% 100 ms,
100% 10 ms, 100% 5000 ms
Perf Criteria A, A, A, B high line,
A, B, A, B low line. See note 5
• ANSI AMMI ES60601-1:2005, CSA C22-2
No 60601-1:08, EN60601-1:2006
Safety Approvals

SDU Related Products

SDU SDU150PS12B SDU150PS24B
Description Medical Approvals Medical Approvals Medical Approvals

Technical ResourceMore

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Robot
development
community

Index Files: 1785  1042  753  1589  1426  36  21  16  32  29 
Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2026 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号