EEWORLDEEWORLDEEWORLD

Part Number

Search

TM41-28

Description
30-40 WATT MEDICAL POWER SUPPLIES
File Size214KB,2 Pages
ManufacturerTRUMPower
Websitehttp://www.trumpower.com/
Download Datasheet Compare View All

TM41-28 Overview

30-40 WATT MEDICAL POWER SUPPLIES

TRUMPower
D E SCR IPTION
30-40 WATT MEDICAL POWER SUPPLIES
TM41 SERIES
( LV D )
The TM41 series of compact, open PCB constructed,
AC-DC switching power supplies are capable of
delivering 30 to 40 watts of continuous output power.
They operate at 85 to 264VAC input voltage without
the need of voltage selection. They are ideally suited
for use in medical equipments, safety systems and
monitoring equipments.
RoHS
F EATU R E S
Low safety ground leakage current
100% burn-in
8 5 t o 2 6 4 VA C w i d e i n p u t r a n g e
Input surge current protection
O v e r v o l ta g e p r o t e c t i o n
Overcurrent protection
Open PCB construction
C o m p l i a n t w i t h R o H S r e q u i r e m e n ts
Saf et y Sta n da rd A ppr o va ls:
UL 2601-1
CSA C22.2 No. 601.1
IN PUT S PE C IF ICAT ION S
I n p u t v o l ta g e
Input frequency
Input current
Earth leakage current
8 5 t o 2 6 4 VA C
11 0 t o 2 8 0 V D C
47 to 63Hz
1 . 1 A
r m s
f o r 11 5 VA C
0 . 6 A
r m s
f o r 2 3 0 VA C
9 0 µ A m a x . @ 11 5 VA C , 6 0 H z
1 5 0 µ A m a x . @ 2 3 0 VA C , 5 0 H z
TÜV EN60601-1
GENERAL SPECIFICATIONS
Switching frequency:
42KHz ±5KHz
Efficiency:
70% minimum on single output
model with Vo
12V, 68%
minimum on the others
20 msec minimum at 110VAC
±0.5% maximum at full load
12 amps @ 115VAC or 24 amps
@ 230VAC, at 25ºC cold start
4000VAC from input to output
1500VAC from input to ground
500VAC from output to ground
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Over voltage protection:
See rating chart
See rating chart
1% peak to peak max.
Provided on output; set at
112 – 132% of its nominal
output voltage
The output protected to short
circuit conditions
All outputs ±0.04%/ºC
i
Maximum excursion of 4% or
better on all models,
recovering to 1% of final value
within 500µs after a 25% step
load change
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
Over current protection:
Temperature coefficient:
Transient response:
600,000 hours minimum at full load
at 25ºC ambient, calculated per
MIL-HDBK- 217F
EMC Performance (EN60601-1-2: 2001)
E N 5 5 0 11 :
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
E N 6 1 0 0 0 - 4 - 11 :
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Class B conducted, Class B radiated
Harmonic distortion, Class A and D
Line flicker
ESD, ±8KV air and ±6KV contact
Radiated immunity, 3V/m for 80-2500MHz
Fast transient/burst, ±2KV
Surge, ±1KV diff, ±2KV com.
Conducted immunity, 3Vrms
Magnetic field immunity, 3A/m
Voltage dips, 30% reduction for 500ms,
60% reduction for 100ms and >95%
reduction for 10ms
MTBF:
E N V IR O N M ENTA L SP E C IF ICATI O N S
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
0ºC to +70ºC
-40ºC to +85ºC
5% to 95% non-condensing
Derate from 100% at +50ºC
linearly to 50% at +70ºC
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
Fax: 408-988-6622
email: sales@trumpower.com
Website: www.TRUMPower.com

TM41-28 Related Products

TM41-28 TM41 TM41-15 TM41-5.1 TM41-24 TM41-12 TM41-18
Description 30-40 WATT MEDICAL POWER SUPPLIES 30-40 WATT MEDICAL POWER SUPPLIES 30-40 WATT MEDICAL POWER SUPPLIES 30-40 WATT MEDICAL POWER SUPPLIES 30-40 WATT MEDICAL POWER SUPPLIES 30-40 WATT MEDICAL POWER SUPPLIES 30-40 WATT MEDICAL POWER SUPPLIES
Ask about burning the system to the development board
Hello everyone, when I was burning the NK file into the development board and reached 99%, the hyperterminal showed the following message: OEMVerifyMemory FAILED !OEMVERIFYMEMORY: Invalid image and th...
unicorn06 Embedded System
Show off your Double Eleven shopping spree
I just got the goods for the Double Eleven shopping spree in the past two days. I've been very busy recently. It's so hard to make a living in this world. A pickit3 for 11.11 yuan and an MCP19111 eval...
ljj3166 Talking
Ten important technical issues in enterprise virtualization projects China IDC Circle
Ten important technical issues for enterprise virtualization projects China IDC Circle Using virtualization technology to extract software from hardware and create a flexible and dynamic environment i...
ran96com Embedded System
Questions about the boot mode of Altera SoC
Seeing the following paragraph, I don’t know if it is true or false. Who is interested in testing it? This question is very critical and important, and is closely related to our system design. [font=宋...
chenzhufly FPGA/CPLD
Fuze target characteristics 4mm (65-85GHz) measurement system
A high-isolation magic-T bridge is used as a transmit-receive switch, and amplitude modulation and phase modulation methods are adopted at one end of the magic-T bridge to complete a single-base milli...
JasonYoo Test/Measurement
[I contribute to the XILINX Resource Center] Design tips--Xilinx internal information
Design Tips--Xilinx Internal InformationThis is a list of common mistakes made in design that can make your design unreliable or slow. To improve the performance of your design and increase the reliab...
wanghongyang FPGA/CPLD

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Robot
development
community

Index Files: 465  2617  1070  969  2803  10  53  22  20  57 
Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2026 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号