EEWORLDEEWORLDEEWORLD

Part Number

Search

TMC100-S05

Description
100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
File Size97KB,2 Pages
ManufacturerTRUMPower
Websitehttp://www.trumpower.com/
Download Datasheet Compare View All

TMC100-S05 Overview

100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES

TRUMPower
DESCRIPTION
100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
TMC100 SERIES
The TMC100 series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 100 watts of
continuous output power at convection. They operate at 90-264 VAC
input voltage without the need of voltage selection, and are suited for
medical, information technology and industrial applications. Approval
to both EN 60601-1 and EN 60950-1 safety standards improves
design-in time and reduces end equipment compliance costs.
FEATURES
Medical and ITE approvals
Compact size 2” × 4” × 1.26”
Ultra high power density 10 W/cubic inch
Full range input 90-132 VAC, or 180-264 VAC
Low earth leakage current
Conducted EMI class B
RoHS compliant
SAFETY STANDARD APPROVALS
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
90-132 VAC or 180-264 VAC
47-63 Hz
1.9A (rms) for 100-120 VAC
1.0A (rms) for 200-240 VAC
150 µA max. @ 264 VAC, 63 Hz
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
OUTPUT SPECIFICATIONS
Output voltage/current:
Total output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
See rating chart
100 watts maximum
250 mV peak to peak on 5.0 V model ,
1%
peak to peak on other models.
Provided on output; set at 110-140% of its
nominal output voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% / C maximum
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
o
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
100 KHz
87-90% @ 230 VAC full load.
12 msec minimum at 110 VAC.
±0.2% maximum at full load
40 A @ 115 VAC or 80 A @ 230 VAC, at 25 C
cold start
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
MTBF:
EMC Performance
EN55011 / EN55022:
FCC:
VCC:
Class B conducted, class A radiated
Class B conducted, class A radiated
Class B conducted, class A radiated
Harmonic distortion, class A
Line flicker
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips,
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 5 sec.
Performance criteria A, B, A.
270,000 hours at full load at 25 C ambient
temperature, calculated per MIL-HDBK-217F
o
o
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
Cooling:
-10 C to +70 C
-40 C to +85 C
5% to 95% non-condensing
Derate from 100% at +50 C linearly to
50% at +70 C
Convection cooling
o
o
o
o
o
o
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Phone: 408-988-6616
Fax: 408-988-6622
email: sales@trumpower.com
Website: www.TRUMPower.com

TMC100-S05 Related Products

TMC100-S05 TMC100-S15 TMC100-S18 TMC100-S36 TMC100-S12 TMC100-S24 TMC100-S28 TMC100-S48
Description 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES 100 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
What are some of the principles of mechanical, galvanometer, and laser scanners?
In order to protect the galvanometer, a lens is added. The laser will be refracted when passing through the lens. So does this refraction need to be handled by software? If the lens moves, then the so...
天文望远镜 Industrial Control Electronics
C51 source code collection
C51 Source Program Collection C language source program sharing collection, including many commonly used program codes and explanations...
呱呱 51mcu
Repost: Tiva C LaunchPad and expansion board
[i=s] This post was last edited by Pinghu Qiuyue on 2014-6-3 12:31 [/i] I uploaded it last time, but it was a semi-finished product. This time, the lab people soldered the components on it, and it can...
平湖秋月 Microcontroller MCU
I recently encountered a problem in studying the implementation of cortex-a8 and linux audio driver
[size=4]I have made an MP3 player based on cortex-a8 before, and now I am studying the implementation of cortex-a8 and linux audio driver with my classmates. The two topics are very similar. The forme...
物联网菜鸟 ARM Technology
How can college graduates enter the embedded industry?
In recent years, the employment of Chinese college graduates has always been a prominent problem. Some people joke that college graduates are unemployed. Millions of students graduate from schools eve...
songbo Embedded System
Android data usage surpasses Apple's
Now more and more people like to use Android phones. Zhuoyue Education found that the Android system not only leads the market share in the United States, but also has a great advantage in traffic. Ac...
zhuoyue TI Technology Forum

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Robot
development
community

Index Files: 1547  1280  2004  1005  350  32  26  41  21  8 
Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2026 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号