EEWORLDEEWORLDEEWORLD

Part Number

Search

RSP-160-5

Description
160W Single Output Medical Type
File Size400KB,2 Pages
ManufacturerMEAN WELL
Websitehttp://www.meanwell.com
Download Datasheet Compare View All

RSP-160-5 Overview

160W Single Output Medical Type

160W Single Output Medical Type
Features :
RPS-160
Universal AC input / Full range
Built-in active PFC function
series
Protections: Short circuit / Overload / Over voltage / Over temperature
5"x3" compact size
Free air convection for 110W and 160W with 20.5 CFM forced air
With power good and fail signal output
Built-in remote sense function for 5~15V
No load power consumption under 0.75W by PS-ON control
(G model)
Standby 5V@0.8A with fan, @0.6A without fan (G model)
3 years warranty
RPS G
G: With 5Vsb & no load power
consumption < 0.75 W
Blank: Basic function (without 5Vsb)
160
12
SPECIFICATION
MODEL
DC VOLTAGE
RPS -160-5
5V
RATED CURRENT (20.5CFM)
30A
CURRENT RANGE (convection)
0 ~ 20A
CURRENT RANGE (20.5CFM)
0 ~ 30A
RATED POWER (convection)
Note.7
103W
RATED POWER (20.5CFM)
Note.8
155W
OUTPUT
RIPPLE & NOISE (max.)
Note.2
100mVp-p
VOLTAGE ADJ. RANGE
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE
FREQUENCY RANGE
POWER FACTOR (Typ.)
INPUT
EFFICIENCY (Typ.)
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
LEAKAGE CURRENT
OVERLOAD
OVER VOLTAGE
PROTECTION
OVER TEMPERATURE
CH1:4.5 ~ 5.5V
VOLTAGE TOLERANCE
Note.3
4.0%
0.5%
1.0%
1800ms, 30ms/230VAC
Note.6
90 ~ 264VAC
RPS -160-12
12V
12.9A
0 ~ 9.1A
0 ~ 12.9A
112.2W
159.8W
100mVp-p
CH1:10.8 ~ 13.2V
3.0%
0.5%
1.0%
RPS -160-15
15V
10.3A
0 ~ 7.3A
0 ~ 10.3A
112.5W
159.5W
120mVp-p
CH1:13.5 ~ 16.5V
3.0%
0.5%
1.0%
RPS -160-24
24V
6.5A
0 ~ 4.6A
0 ~ 6.5A
113.4W
161W
150mVp-p
CH1:22 ~ 27V
2.0%
0.5%
1.0%
RPS -160-48
48V
3.25A
0 ~ 2.3A
0 ~ 3.25A
113.4W
161W
250mVp-p
CH1:43.2 ~ 52.8V
2.0%
0.5%
1.0%
3500ms, 30ms/115VAC at full load
16ms/230VAC/115VAC at full load
127 ~ 370VDC
PF>0.98/115VAC at full load
87%
87%
87%
88%
47 ~ 63Hz
PF>0.93/230VAC
85%
2A/115VAC
1.1A/230VAC
COLD START 35A/115VAC
70A/230VAC
Earth leakage current <300uA / 264VAC, patient leakage current <100uA/264VAC
105 ~ 135% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
5.75 ~ 6.75V
13.8 ~ 16.2V
17.25 ~ 20.25V
27.6 ~ 32.4V
55.2 ~ 64.8V
Protection type : Shut down o/p voltage, re-power on to recover
110 (5V),105 (12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor)
95
(5V),90
(12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor)
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover
5VSB : 5V@0.6A without fan, 0.8A with fan 20.5CFM ; tolerance 2%, ripple : 50mVp-p(max.)
5V STANDBY (G model)
FUNCTION
PS-ON INPUT SIGNAL (G model)
Power on: PS-ON = "Hi" or " > 2 ~ 5V" ; Power off: PS-ON = "Low" or " < 0 ~ 0.5V"
PF>1ms
POWER GOOD / POWER FAIL
500ms>PG>10ms
5 ~ 15V
REMOTE SENSE
-20 ~ +70
(Refer to "Derating Curve")
WORKING TEMP.
WORKING HUMIDITY
ENVIRONMENT STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
SAFETY STANDARDS
SAFETY & WITHSTAND VOLTAGE
ISOLATION RESISTANCE
EMC
(Note 4)
EMC EMISSION
20 ~ 90% RH non-condensing
-40 ~ +85 , 10 ~ 95% RH
0.03%/ (0 ~ 50 )
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1,
TUV EN60601-1 approved
I/P-O/P:4KVAC
I/P-FG:1.5KVAC
O/P-FG:0.5KVAC
I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A
230.5Khrs min. MIL-HDBK-217F (25 )
127*76.2*34.6mm (L*W*H)
0.32Kg; 36pcs/12.5Kg/0.79CUFT
EMC IMMUNITY
MTBF
OTHERS
DIMENSION
PACKING
NOTE
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
5. HS1,HS2 & HS3 can not be shorted.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. The rated power includes 5Vsb @ 0.6A.
8. The rated power includes 5Vsb @ 0.8A.

RSP-160-5 Related Products

RSP-160-5 RPS-160_12 RSP-160-12 RSP-160-15 RSP-160-24 RSP-160-48
Description 160W Single Output Medical Type 160W Single Output Medical Type 160W Single Output Medical Type 160W Single Output Medical Type 160W Single Output Medical Type 160W Single Output Medical Type
MSP430F149 minimum system board uses 1602 LCD display sine wave oscilloscope program control
First, let's talk about the structure of MSP430F149. MSP430 series microcontrollers are 16-bit ultra-low power mixed signal processors with precise instruction set (RISC) that Texas Instruments (TI) b...
fish001 Microcontroller MCU
AD7280 Register Configuration
If only one device is used, how to set the control register to initialize the device address? What instructions should be issued? The instructions provided by the official routine (daisy chain mode): ...
gaorz_ ADI Reference Circuit
Please recommend a serial flash
I need a 1Mbyte serial flash (non-IIC interface). I have ST's 25P80 and ATMEL's AT45DB081. I don't know which one is cheaper or if there is a cheaper 1Mbyte serial flash. My boss is very cost-consciou...
Maywolf stm32/stm8
Urgent!!! Please help. Thank you very much. Waiting online
I am a beginner in ucosii. I have a thesis defense tomorrow. The thesis is about the kernel porting of ucosii. I have a development board. Please explain the porting process in detail. My phone number...
pusongling Embedded System
WINCE system freezes after power on
The WINCE system crashes irregularly after being powered on. Could you please give me some ideas on how to find the problem? Sometimes the system crashes while running. Could you please give me some i...
danielyang Embedded System
Warmly call for sharing happiness!
It is a success for forum members to share their work problems and gains, and it is also a help to others! To put it in a higher way, it is to make the motherland stronger!...
eeleader Industrial Control Electronics

EEWorld
subscription
account

EEWorld
service
account

Automotive
development
circle

Robot
development
community

Index Files: 1756  101  1510  2407  1760  36  3  31  49  38 
Datasheet   0 1 2 3 4 5 6 7 8 9 A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
Room 1530, 15th Floor, Building B, No. 18 Zhongguancun Street, Haidian District, Beijing Telephone: (010) 82350740 Postal Code: 100190
Copyright © 2005-2026 EEWORLD.com.cn, Inc. All rights reserved 京ICP证060456号 京ICP备10001474号-1 电信业务审批[2006]字第258号函 京公网安备 11010802033920号